Buy Plasma Freezer in Nigeria.
Buy a plasma freezer in Nigeria through Franance by matching the exact model's documented temperature range, capacity, monitoring and electrical variant to your blood service's written storage and processing requirements. The current category lists Haier Biomedical DW-30L528D and Thermo Scientific TSX2330LV. Their published ranges and configurations differ; in particular, Haier documents DW-30L528D as a general -30°C biomedical freezer, so its fit for a plasma protocol must be confirmed in writing.
Products
2 models
Where can I buy a plasma freezer in Nigeria?
Browse the plasma freezer category. It contains two currently assigned products: the Haier Biomedical DW-30L528D listing and the Thermo Scientific TSX2330LV listing. Each exact model filter returns one result in the current catalogue.
Read moreWhich plasma freezers are listed?
Franance currently assigns two products to medical-refrigeration/blood-bank-refrigeration/plasma-freezer:
What is the Haier Biomedical DW-30L528D specification?
Haier's DW-30L528D technical data sheet identifies an upright 531 L / 18.8 cu ft direct-cooling freezer. It lists manual defrost, ten drawers, -30°C cooling performance and a temperature range from -10°C to -30°C. The sheet documents two power variants, 220–240 V/50 Hz and 115 V/60 Hz, as well as high/low-temperature, remote, power-failure, sensor-error, high-ambient-temperature and door-ajar alarms. It lists USB and RS485 as optional interfaces and the temperature recorder as optional.
Read moreBuy plasma freezers in Nigeria
Buy a plasma freezer in Nigeria through Franance by matching the exact model's documented temperature range, storage capacity, monitoring and electrical configuration to the blood service's written requirements. The category currently lists two products: Haier Biomedical DW-30L528D and Thermo Scientific TSX2330LV. Their published configurations differ. Haier identifies DW-30L528D as a general -30°C biomedical freezer, so its suitability for a specific plasma storage protocol must be confirmed with Haier in writing before procurement. A freezer's category label, nominal cooling performance or setpoint alone does not establish compliance with a facility's procedure or validated performance under load.
Where can I buy a plasma freezer in Nigeria?
Browse the plasma freezer category. It contains two currently assigned products: the Haier Biomedical DW-30L528D listing and the Thermo Scientific TSX2330LV listing. Each exact model filter returns one result in the current catalogue.
Franance handles these requests as quote orders. Send the model you want reviewed, the intended plasma product and storage protocol, required capacity and temperature range, facility location, site power details, monitoring needs and quantity. Ask the quotation to state the offered model and electrical variant, accessories, delivery, installation or commissioning scope, warranty territory and support route. The product listing does not confirm stock, price, delivery date or after-sales service in Nigeria; those details need confirmation for the actual order.
Which plasma freezers are listed?
Franance currently assigns two products to medical-refrigeration/blood-bank-refrigeration/plasma-freezer:
- Haier Biomedical DW-30L528D is an upright, direct-cooling biomedical freezer with a 531 L (18.8 cu ft) capacity. Haier's model sheet lists ten drawers, a -30°C cooling performance and a -10°C to -30°C operating range. It lists 220–240 V/50 Hz and 115 V/60 Hz power variants. Haier's technical sheet calls it a biomedical freezer; it does not itself document a plasma-specific storage protocol or confirm that the freezer meets a particular blood service's required temperature range.
- Thermo Scientific TSX2330LV is listed by Thermo Fisher as a TSX Series plasma freezer with 650 L (23 cu ft) capacity. The LV variant is listed with a CEE 7/7 plug, 208–230 V at 50/60 Hz, a standard chart recorder, seven stainless-steel drawers, an adjustable temperature range from -35°C to -15°C and a -30°C setpoint. Thermo's specification also lists automatic defrost and dimensions; confirm current details for the actual offered unit.
The exact DW-30L528D and TSX2330LV listings isolate the current products. These model facts come from manufacturer documents. Compare current revision, local electrical version and included accessories on the written quotation rather than relying only on the shortened product title.
What is the Haier Biomedical DW-30L528D specification?
Haier's DW-30L528D technical data sheet identifies an upright 531 L / 18.8 cu ft direct-cooling freezer. It lists manual defrost, ten drawers, -30°C cooling performance and a temperature range from -10°C to -30°C. The sheet documents two power variants, 220–240 V/50 Hz and 115 V/60 Hz, as well as high/low-temperature, remote, power-failure, sensor-error, high-ambient-temperature and door-ajar alarms. It lists USB and RS485 as optional interfaces and the temperature recorder as optional.
One point requires special care: the local catalogue files the unit under plasma freezers, but Haier's model-specific sheet describes a general biomedical freezer. The model's lowest listed temperature is -30°C. Neither the category assignment nor the manufacturer's general freezer family description establishes suitability for a particular blood service's plasma product, processing method, storage duration, alarm criteria or applicable protocol. Before comparing it with a plasma freezer specifically marketed for that use, give Haier the facility's written requirements and request a response that names DW-30L528D and the exact electrical/configuration variant.
Capacity is also more than the gross 531 L figure. Check usable shelf or drawer space, bag or box dimensions, the normal number of units in storage, peak inventory and how staff organize and retrieve units. Haier's technical sheet lists ten drawers but does not establish that every bag, carton or local inventory arrangement will fit in a way that meets operating procedures. Ask for internal layout details and a loading plan before order.
What is the Thermo Scientific TSX2330LV specification?
Thermo Fisher's TSX2330LV specification page identifies a 650 L / 23 cu ft plasma freezer. The LV product number is listed with a CEE 7/7 plug and 208–230 V/50–60 Hz supply. Thermo lists a standard chart recorder, seven stainless-steel drawers, automatic defrost, a -35°C to -15°C range and a -30°C setpoint. The page includes interior and exterior dimensions and lists compatible wireless monitoring options. Verify the exact current version and any selected monitoring accessories on the quote.
Thermo describes the TSX Series as plasma freezers and publishes product-level compliance information for the line. Do not convert that manufacturer statement into a blanket claim of Nigerian approval or compliance with a facility's blood service requirements. The quote should include the exact model, current product documentation, recorder and monitoring scope, electrical specification, plug, dimensions, included drawers and any optional monitoring devices. Thermo states that warranty varies by country, so obtain the warranty territory and process in writing for the Nigerian destination.
The model page shows the LV variant alongside other plug and voltage variants. The TSX2330LV filter is specific to the current catalog item; choosing the right name is not enough if a supplier offers a different regional suffix or electrical setup. Ask the supplier to confirm the catalog number on the order document and on the delivered unit's rating plate.
Are the DW-30L528D and TSX2330LV interchangeable?
They should not be treated as interchangeable solely because both are categorized as plasma freezers. The Thermo TSX2330LV is explicitly identified by its manufacturer as a plasma freezer and has a published -35°C to -15°C range with a -30°C setpoint. Haier DW-30L528D is a general biomedical freezer whose sheet lists a -10°C to -30°C range and does not identify a plasma-specific operating protocol. Their capacities, drawers, alarm/recording arrangements, electrical variants and documentation also differ.
Make the comparison against the written requirements from the blood bank or transfusion service: what plasma type is held, its processing and storage workflow, the required operating range and alarm response, the expected inventory, documentation and quality-system controls. The responsible service should determine the requirements; the vendor should confirm the precise model can meet them. If the Haier model remains under consideration, ask for an explicit, model-specific statement of fit against the protocol instead of inferring suitability from a general -30°C class or catalogue taxonomy.
WHO's blood cold-chain guide is a procurement reference covering the selection of equipment and accessories for blood storage and transport. It discusses factors such as capacity, power, temperature monitoring, ambient conditions, servicing and the intended cold-chain process. It does not validate a specific contemporary commercial model or replace current national procedures. Have the quality lead and biomedical/clinical engineering team review current applicable guidance and model documentation together.
How should I calculate freezer capacity for plasma storage?
Estimate usable capacity from the real inventory, not just nominal litres. Count routine and peak stored units, the package or box dimensions, how units are grouped, the number of locations required for access and any reserve space needed for planned growth. Include the normal receiving and retrieval pattern. Ask the manufacturer for the internal dimensions, drawer usable dimensions and box capacity for the exact model. Thermo lists a 650 L cabinet with seven drawers and gives a box-holding figure on its product page; Haier lists 531 L and ten drawers. Those numbers describe the products but do not prove that your preferred loading layout fits.
Do not pack beyond the manufacturer's loading instructions or facility procedure. Ask what clearances are needed around stored bags or boxes to preserve the specified airflow and access. Confirm if a full load affects pull-down or recovery claims, and whether the manufacturer has data for your intended load. The quoted capacity should state whether it is gross volume, usable volume, number of standard boxes or another measure. Use the actual container type and inventory profile when assessing the result.
If several departments will share equipment, clarify ownership, access, segregation, inventory tracking and backup arrangements before selecting a cabinet. A freezer that physically holds the projected quantity may still be a poor operational fit if staff cannot identify and retrieve items under the facility's procedures. A site or workflow review can resolve these questions before the purchase order.
What temperature monitoring and records should I specify?
Begin with the blood service's written limits, alarm response procedures and records policy. Specify the required setpoint and acceptable range, alarm thresholds and delay, local audible/visual indication, remote alarm contacts, data logging or chart records, data retrieval, calibration, review frequency and retention responsibilities. Confirm whether the monitor, recorder, sensor, network gateway, software or license is included, optional or supplied by the facility.
The two models differ in the information published by their manufacturers. Thermo lists a standard chart recorder and wireless monitoring compatibility for TSX2330LV. Haier lists high/low temperature, remote, power failure, sensor-error, high-ambient and door-ajar alarms for DW-30L528D; USB/RS485 interfaces and the temperature recorder are listed as optional. Confirm the offered option codes and exact interface behavior with the manufacturer. A remote-alarm terminal is not the same as an installed monitoring service, and compatibility does not mean a monitoring system is included.
Ask how alarms are relayed, what happens during a network interruption, who receives escalation, how sensor performance is checked and what evidence is provided at handover. Clarify whether site mapping, commissioning, qualification or performance testing is available and who is responsible. Do not assume that a product brochure or factory specification proves the room installation meets the facility's monitoring and validation plan.
What power and site requirements should I check in Nigeria?
Confirm the actual supply where the freezer will operate before ordering. Record voltage, frequency, available circuit capacity, protective devices, socket and plug type, power quality, earthing and planned backup power with qualified facility engineering staff. The DW-30L528D sheet lists 220–240 V/50 Hz and a separate 115 V/60 Hz variant. The Thermo TSX2330LV is listed at 208–230 V/50–60 Hz with a CEE 7/7 plug. A product listing does not establish which variant will be supplied or whether an adapter, stabilizer or other electrical component is appropriate for the site.
Measure the route from delivery point to room, including lift and doorway clearances, corridor turns, thresholds and floor-loading restrictions. Thermo lists exterior dimensions of about 96 × 71.1 × 199.6 cm for TSX2330LV and a unit weight of about 200 kg. Haier lists exterior dimensions of about 940 × 845 × 1878 mm and net weight of about 180 kg for DW-30L528D. Check door swing, service clearance, ventilation, heat rejection, cleaning access and condenser access against each current installation manual.
Discuss ambient temperature and ventilation with the manufacturer. Haier lists climate-class and high-ambient alarm information for the exact model; this should not be read as permission to install it in any hot or poorly ventilated room. Ask the supplier for the room limits, clearance diagram and heat output for the offered variant. In Nigeria, have facilities engineering verify the installation conditions and power continuity plan. Do not assume a freezer eliminates the need for backup procedures during an outage.
How should the freezer fit into a blood cold-chain workflow?
A freezer is one piece of a controlled process covering receipt, identification, storage, monitoring, issue, transport, contingency and maintenance. The World Health Organization describes the blood cold chain as a system for storing and transporting blood products under the required conditions. Its equipment procurement guide and equipment management manual help teams consider equipment selection, operating procedures and maintenance needs. Use current national and facility rules for the actual product and workflow; this guide is not clinical or transfusion advice.
Before purchase, establish how the freezer's locations, drawers and locks support the service's identification and access control procedures. Decide who logs new inventory, reviews alarms, authorizes removal, handles excursions and contacts maintenance. If plasma will be transported from or to another location, specify transport equipment separately and assess the complete route; a static storage freezer does not provide a validated transport solution.
Plan preventative maintenance, cleaning, alarm checks, temperature review, spare parts and repair escalation. Ask what service documentation, user training and maintenance schedule the manufacturer recommends for the exact model. Confirm who can provide service in the destination region and the response process in writing. No local service route is established by the product page alone.
What should a plasma freezer quotation include in Nigeria?
For a useful quotation, provide the exact model or approved alternative, the responsible blood-service storage requirement, required temperature range and setpoint, current and projected inventory, unit and box dimensions, loading pattern, alarms and records requirements, and any remote monitoring or interface need. Name the facility, city and state, requested quantity, intended installation room, measured voltage/frequency, plug and circuit details, room ambient conditions, available backup power and delivery access.
Ask Franance and the manufacturer to identify the exact catalog number, model revision and electrical variant. Request current datasheet, installation/operation manual and model-specific confirmation against the stated plasma protocol, particularly for DW-30L528D. Ask the quote to show included drawers, chart recorder, sensors, alarms, cables, accessories, delivery, installation, commissioning or qualification, user training, maintenance, warranty and service as separate line items. Mark optional or facility-supplied items clearly.
Confirm price and currency, validity period, stock status, lead time, export/import responsibilities, freight, insurance, customs and delivery terms for Nigeria. Obtain written details for the warranty territory, authorized service path, spare-part availability and response arrangements. Do not assume the manufacturer's global page establishes Nigerian inventory, approval, warranty or service coverage. The facility should accept the selected model only after its responsible technical and quality teams have reviewed the exact documentation and implementation scope.
How do I place a plasma freezer quote order?
Open the Haier DW-30L528D listing or the Thermo Scientific TSX2330LV listing, then send the model, storage requirement, capacity plan, site power, monitoring needs, quantity and Nigerian destination through a plasma freezer quotation request. Franance will follow up by phone or email to clarify the quote order. There is no online checkout.
Manufacturer and WHO sources; scope limits
The model-specific facts in this guide come from Haier Biomedical's DW-30L528D technical sheet and biomedical freezer range page, and Thermo Fisher's TSX2330LV specification page and TSX plasma freezer installation and operation manual. WHO sources are the Blood cold chain: equipment selection and procurement guide, the blood cold-chain overview and the blood cold-chain equipment manual.
Haier's local catalogue category assignment does not replace its model-specific documentation. The DW-30L528D guide therefore flags its general biomedical freezer classification and requires confirmation against a plasma protocol. Thermo publishes a model explicitly named a plasma freezer; its published specification still does not validate performance at a particular loaded site. This guide does not certify either unit, determine the correct clinical storage protocol or assert Nigerian availability, price, local regulatory approval, installation, warranty or service. Confirm current documents, exact supplied variants and all commercial terms with the manufacturers, Franance and the responsible facility teams.
Common questions
Does DW-30L528D meet my blood service's plasma storage requirement?
The model sheet documents a general biomedical freezer, -30°C cooling performance and -10°C to -30°C range. That information alone does not establish suitability for a specific plasma protocol. Give Haier the required protocol and obtain written confirmation for the exact model and variant.
What does the TSX2330LV setpoint mean?
Thermo Fisher lists an adjustable range from -35°C to -15°C and a -30°C setpoint for TSX2330LV. Confirm the current model documentation, site conditions and facility-required performance; the setpoint is not proof of loaded cabinet performance or validation at your site.
Are these freezers ready to operate on any Nigerian power supply?
Do not assume that. DW-30L528D is documented with 220–240 V/50 Hz or 115 V/60 Hz variants; TSX2330LV is listed as 208–230 V/50–60 Hz with CEE 7/7 plug. Confirm the supplied variant, measured site voltage/frequency, circuit and plug arrangement with the supplier and qualified facility engineering staff.
Can Franance confirm local stock or service from the listing?
No. Availability, price, lead time, delivery, warranty territory, installation and support must be confirmed for the exact quote and Nigerian destination.
Request a plasma freezer quotation in Nigeria.
Send Franance the exact model, written storage protocol, required range and capacity, loading plan, site power and monitoring needs, destination, installation/qualification request, maintenance expectations and quantity. Ask the manufacturer and supplier to confirm the offered configuration against the blood service requirement in writing.
Request a plasma freezer quoteNeed an equipment quote?
Add the models you need to a quote order, or send us your equipment list.